About ROSAT |
ROSAT Home Page | ROSAT Images |
---|
G. Fraser [Fraser1992] has analyzed the on-axis ground based data by fitting Pearson functions to the 7 pulse height distributions shown in Fig. 5.16. A detailed analysis of his results are given in his report. The relative FWHM of the fitted pulse height distributions vary from 66% (for the Al-K and Fe-K lines) to 107%(for the B-K line). Fraser concludes that ``the average pulse height per pixel has very little dynamic range and is a very poor indicator of photon energy''. However, he also notes that ``the softness ratio, (counts in channels 1-5)/(counts in channels 6-11), used by [Wilson et al.1992] does seem the most sensitive energy indicator that can be constructed.'' (As noted below, the actual channels fulfilling this purpose will change as a function of the temporally-dependent gain.)
Figure 5.17: The spatial variation in the centroid of the pulse height
distribution in the ground based Cu(0.93 keV) flat field image.
The image was first blocked by a factor of 64, producing 32''
pixels with 1% statistical errors. The contours are at PHA values
of 4.0, 4.25, 4.5, 5.0, 5.5, 6.0, 7.0 and increase outward.
Figure 5.18: PVC phase observations of HZ 43 and AR Lac were taken at many
positions on the detector surface. The observed mean PHA of
the calibration source measurements are plotted with errors
against the expected mean PHA from the ground based
measurements of copper (0.93 keV). The line represents the
best least squares fit of the data points.
Figure 5.19: The pulse height distribution in 3 HRI observations of N132D.
The highest gain observation (solid line) was taken during
the PVC phase, the lowest gain observation was taken in
1993 April, and the observation shown in the dot-dash line
was taken in 1994 July, just after the HV increase.
Also shown (dashed lines) are the pulse height distributions
of the Al-K (1.49 keV) and B-K (0.18 keV) lines obtained
during the ground based tests.
Although the total count rate (in all PHA channels) has
remained nearly constant
(see Tab. 5.4),
the earlier observations show a downward trend in the peak
of the distribution. The latest observation shows that the
HV increase has restored the peak to within 0.5 channels
of the original.
Using the ground-based data (see Fig. 5.16) we constructed a spectral redistribution matrix for the HRI. We then fit the pulse height distributions of several observations of HZ 43 and one observation of 3C273. PSPC spectra of HZ 43 are well fit with an unabsorbed black body spectrum with a temperature of 0.02keV ( for 24 degrees of freedom). Table 5.8 lists the values obtained by fitting the HRI observations of HZ 43 with the same unabsorbed black body spectrum, with only the normalization treated as a free parameter. The minimum values are extremely high no matter what PHA channels are included in the fit. The best fit absorbed power-law model to the PSPC spectrum of 3C273 is obtained with cm and an energy index of . Since the HRI observation of 3C273 has 30,000 net counts we fit the HRI spectra with 0, 2, and 5% systematic errors, added in quadrature with the statistical errors. Table 5.9 lists the minimum values obtained by fitting the HRI\ observation of 3C273 with and fixed. We also fit the HRI pulse height distribution treating as a free parameter. The best fit energy index so obtained for 3C273 is significantly steeper than that obtained from the PSPC observation.